Contact AFM
Basic Contact Mode AFM Operation
124 Dimension 3100 Manual Rev. D
Feedback Controls Panel
1. Set the Integral gain to 2.0 and the Proportional gain to 4.0 (see Figure 8.1e).
2. Set the Deflection Setpoint to 0V.
Figure 8.1e Suggested Feedback Controls Settings
Channel Panels 1 and 2
1. In the Channel 1 panel, set Data type to Height (see Figure 8.1f).
2. Set Data scale to a reasonable value for the sample.
Note: For example, for a 200nm step height calibration sample, a reasonable Data
scale setting is 300nm initially.
3. Set Line direction to either Trace or Retrace.
4. On the Channel 2 panel, verify Data type is set to Off to disable the panel.
Figure 8.1f Suggested Channel Controls Settings