Page 1 - Dimension 3100 Manual

Document Revision History: DImension 3100 Manual Revision Date Section(s) AffectedRef. DCRApproval D 03/09/04 All. N/A C. KowalskiC 05/15/03 Fluid Ce

Page 2

x Dimension 3100 Manual Rev. D15.4 Installation of the Electronics Modules . . . . . . . . . . . . . . . . . . . . . . . 26815.4.1 Phase Extender

Page 3 - Ta ble of Contents

Stage SystemStage Menu Commands80 Dimension 3100 Manual Rev. DTeaching a Programmed MoveTo program (teach) a series of moves, complete the following:1

Page 4

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 817. Click on OK in the Teach Mode panel when the stage has moved to a desired position (s

Page 5

Stage SystemStage Menu Commands82 Dimension 3100 Manual Rev. DTo insert a step into an existing program sequence:6. Go to the Stage pop-down menu to t

Page 6

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 833. Select the Run option. If you are loading and running the program from the beginning,

Page 7

Stage SystemStage Menu Commands84 Dimension 3100 Manual Rev. DFigure 5.2p Stage Initialize/Cancel Prompt3. To begin initialization, click Initialize

Page 8

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 85/(86 Blank)7. Finally, the camera’s optics zooms out. When the camera zooms out to the l

Page 10

Rev. D Dimension 3100 Manual 87Chapter 6 Cantilever PreparationThe Dimension 3100 Scanning Probe Microscope comes furnished with etched silicon cant

Page 11

Cantilever PreparationSilicon Cantilever Substrates88 Dimension 3100 Manual Rev. D6.1 Silicon Cantilever Substrates 6.1.1 Wafer Tool KitA wafer tool k

Page 12

Cantilever PreparationSilicon Cantilever SubstratesRev. D Dimension 3100 Manual 89Note: Cover the petri dish when not in use.Figure 6.1a Silicon Can

Page 13 - List of Figures

Rev. D Dimension 3100 Manual xi17.3 Check Sample Orthogonality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31417.3.1 Measure Ort

Page 14

Cantilever PreparationSilicon Cantilever Substrates90 Dimension 3100 Manual Rev. D6.1.3 Tip Shape of Etched Silicon ProbesEtched silicon probes provid

Page 15

Cantilever PreparationSilicon Cantilever SubstratesRev. D Dimension 3100 Manual 91Figure 6.1c Silicon Probe Tip Profile Artifact (Front-to-Back)To me

Page 16

Cantilever PreparationSilicon Cantilever Substrates92 Dimension 3100 Manual Rev. DFigure 6.1d Silicon Probe Tip Step Profile Artifact (Side-to-Side)M

Page 17

Cantilever PreparationSilicon Nitride Cantilever SubstratesRev. D Dimension 3100 Manual 93In addition to microscopic scale shape characteristics (see

Page 18

Cantilever PreparationSilicon Nitride Cantilever Substrates94 Dimension 3100 Manual Rev. D5. Place the strip down on a white piece of paper to inspect

Page 19

Cantilever PreparationSilicon Nitride Cantilever SubstratesRev. D Dimension 3100 Manual 95Figure 6.2b Substrate Break-offNote: Extra substrates are

Page 20

Cantilever PreparationSilicon Nitride Cantilever Substrates96 Dimension 3100 Manual Rev. D6.2.1 Tip Shape of Silicon Nitride ProbesSilicon nitride pro

Page 21 - Chapter 1 System Overview

Cantilever PreparationSilicon Nitride Cantilever SubstratesRev. D Dimension 3100 Manual 97/(96 Blank)Figure 6.2d Silicon Nitride Cantilevers (profile

Page 23 - 1.2 Control Station Overview

Rev. D Dimension 3100 Manual 99Chapter 7 Head, Probe, & Sample PreparationThis chapter includes information regarding the Dimension 3100 Scannin

Page 24 - 1.2.2 Computer

xii Dimension 3100 Manual Rev. D

Page 25 - Control Station Overview

Head, Probe, & Sample PreparationSystem Information100 Dimension 3100 Manual Rev. D• Load the Sample: Section 7.2.12• Focus Surface: Section 7.2.1

Page 26

Head, Probe, & Sample PreparationSystem InformationRev. D Dimension 3100 Manual 1017.1.3 Laser RequirementsThis instrument uses a semiconductor la

Page 27 - Dimension 3100 SPM Overview

Head, Probe, & Sample PreparationBasic AFM Operation102 Dimension 3100 Manual Rev. D7.2 Basic AFM Operation7.2.1 Select the Microscope1. Set the m

Page 28

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 103Figure 7.2b Cantilever Holder Stand (top view)Note: You may

Page 29 - 1.3.4 Dimension SPM Head

Head, Probe, & Sample PreparationBasic AFM Operation104 Dimension 3100 Manual Rev. DFigure 7.2c Standard AFM Cantilever Holder4. Orient the cant

Page 30

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 1057.2.5 Remove the Dimension SPM Head1. Tighten the screw locate

Page 31

Head, Probe, & Sample PreparationBasic AFM Operation106 Dimension 3100 Manual Rev. D1. Mate the cantilever holder sockets to the Dimension SPM hea

Page 32

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 107Figure 7.2e Dimension Head Laser Control KnobsThe X directio

Page 33 - 1.3.5 Cantilever Holder

Head, Probe, & Sample PreparationBasic AFM Operation108 Dimension 3100 Manual Rev. D2. Verify the laser beam is visible on the surface below. If i

Page 34 - 1.3.6 Video Zoom Microscope

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 109Figure 7.2g Photodetector Adjustment KnobsNote: You may now

Page 35 - Sample Size & Handling

Rev. D Dimension 3100 Manual xiiiList of FiguresChapter 1 System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 36 - 1.6 Applications

Head, Probe, & Sample PreparationBasic AFM Operation110 Dimension 3100 Manual Rev. D3. Turn the rear-right laser control knob clockwise to move th

Page 37 - Chapter 2 Safety

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 111Note: During this step do not adjust the back-right laser knob

Page 38

Head, Probe, & Sample PreparationBasic AFM Operation112 Dimension 3100 Manual Rev. DFigure 7.2i Vision System Window3. Center the laser detector

Page 39 - 2.1 Safety Requirements

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 113Figure 7.2j Photodetector Mirror Adjustment Knobs7.2.11 Loca

Page 40 - 2.2 Safety Precautions

Head, Probe, & Sample PreparationBasic AFM Operation114 Dimension 3100 Manual Rev. D7.2.12 Load the SampleIf it is your first time operating the mi

Page 41 - Safety Precautions

Head, Probe, & Sample PreparationBasic AFM OperationRev. D Dimension 3100 Manual 115Figure 7.2l Securing the Sample5. Secure the sample atop the

Page 42

Head, Probe, & Sample PreparationBasic AFM Operation116 Dimension 3100 Manual Rev. D7.2.14 Cantilever Tune (TappingMode only)1. Select Cantilever

Page 43 - 2.2.2 Microscope

Head, Probe, & Sample PreparationAdvanced AFM OperationRev. D Dimension 3100 Manual 1177.2.16 Engage• Select Motor > Engage, or click the Engag

Page 44

Head, Probe, & Sample PreparationAdvanced AFM Operation118 Dimension 3100 Manual Rev. DFigure 7.3a Default SPM Stage ParametersDefinitions of Sta

Page 45

Rev. D Dimension 3100 Manual 119Chapter 8 Contact AFMThis chapter covers procedures for operating the Dimension 3100 Scanning Probe Microscope (SPM)

Page 46 - 2.3 Ergonomics

List of Figuresxiv Dimension 3100 Manual Rev. DFigure 3.3c IS3K-2 - Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Page 47 - 2.4 Environmental Conditions

Contact AFMBasic Contact Mode AFM Operation120 Dimension 3100 Manual Rev. D• Lowpass Filter: Section 8.3.5• Highpass Filter: Section 8.3.6• Force Cali

Page 48 - 2.6.1 Pre Power-up Checklist

Contact AFMBasic Contact Mode AFM OperationRev. D Dimension 3100 Manual 1218.1.5 Adjust Photodetector1. Adjust the photodetector so that the red dot m

Page 49 - Connections

Contact AFMBasic Contact Mode AFM Operation122 Dimension 3100 Manual Rev. D8.1.8 Show All ItemsBefore changing any parameters, you should display all

Page 50 - Final Installation

Contact AFMBasic Contact Mode AFM OperationRev. D Dimension 3100 Manual 1238.1.9 Set Initial Scan ParametersScan Controls PanelIn the Scan Controls pa

Page 51 - Power-up (Installation Only)

Contact AFMBasic Contact Mode AFM Operation124 Dimension 3100 Manual Rev. DFeedback Controls Panel1. Set the Integral gain to 2.0 and the Proportional

Page 52 - 2.8 Software Power-up

Contact AFMAdvanced Atomic Force OperationRev. D Dimension 3100 Manual 1258.1.10 Engage1. Select Motor > Engage. A pre-engage check begins, followe

Page 53 - 2.8.2 Log On

Contact AFMAdvanced Atomic Force Operation126 Dimension 3100 Manual Rev. Dthickness of both cantilevers are equal. The calculated spring-constant for

Page 54 - 2.8.4 Select Realtime

Contact AFMOptimization of Scanning ParametersRev. D Dimension 3100 Manual 127Note: The smaller end radius of etched silicon tips creates greater forc

Page 55 - 2.9 Hazard Labels

Contact AFMOptimization of Scanning Parameters128 Dimension 3100 Manual Rev. DCollect Deflection data with low feedback gains to ensure the piezo remai

Page 56 - 2.9.1 Laser Warning Labels

Contact AFMOptimization of Scanning ParametersRev. D Dimension 3100 Manual 1298.3.3 Scan Size and Scan RateIn general, decrease the Scan rate as the S

Page 57

List of FiguresRev. D Dimension 3100 Manual xvFigure 5.2r Stage Initializing Prompt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84Fig

Page 58 - 3.1 Optional Configurations

Contact AFMForce Calibration Mode130 Dimension 3100 Manual Rev. D8.4 Force Calibration ModeThe Force Calibration command in the View > Force Mode &

Page 59 - 3.1.2 VT-102

Rev. D Dimension 3100 Manual 131Chapter 9 TappingMode AFMThis chapter details procedures for operating the Dimension 3100 SPM in TappingMode in air.

Page 60 - 3.1.3 IS3K-2

TappingMode AFMPrinciples of TappingMode132 Dimension 3100 Manual Rev. D• Decreasing the Cantilever Drive Frequency: Section 9.4.3• Optimization of Sc

Page 61 - 3.2 Facilities Requirements

TappingMode AFMBasic TappingMode AFM OperationRev. D Dimension 3100 Manual 133Figure 9.1b represents the same cantilever at the sample surface. Althou

Page 62 - Supply Module

TappingMode AFMBasic TappingMode AFM Operation134 Dimension 3100 Manual Rev. D9.2.2 Head, Cantilever and Sample Preparation• Install an etched single

Page 63 - Facilities Requirements

TappingMode AFMBasic TappingMode AFM OperationRev. D Dimension 3100 Manual 1359.2.6 Focus Surface1. Select Stage > Focus Surface or click the Focus

Page 64

TappingMode AFMBasic TappingMode AFM Operation136 Dimension 3100 Manual Rev. DAutomatic TuningFor most purposes, the Auto Tune function will suffice.•V

Page 65

TappingMode AFMBasic TappingMode AFM OperationRev. D Dimension 3100 Manual 137Note: The Offset command sets the center frequency equal to the cursor p

Page 66 - Side and Front View

TappingMode AFMBasic TappingMode AFM Operation138 Dimension 3100 Manual Rev. DFigure 9.2c Cantilever Tune Frequency Sweep11. Click OK. The parameter

Page 67 - Front View

TappingMode AFMBasic TappingMode AFM OperationRev. D Dimension 3100 Manual 1399.2.8 Show All ItemsBefore changing any parameters, you should display a

Page 68 - 3.3.5 ELCON Console

List of Figuresxvi Dimension 3100 Manual Rev. DFigure 9.2a Auto Tune Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136Figu

Page 69

TappingMode AFMBasic TappingMode AFM Operation140 Dimension 3100 Manual Rev. D9.2.9 Set Initial Scan ParametersScan Controls PanelIn the Scan Controls

Page 70 - 1959-940-A

TappingMode AFMBasic TappingMode AFM OperationRev. D Dimension 3100 Manual 141Feedback Controls Panel• Set the Integral gain to 0.5 and the Proportion

Page 71 - General Facilities Guidelines

TappingMode AFMWithdraw the Tip142 Dimension 3100 Manual Rev. D9.2.11 Optimize Scan ParametersThe procedure to optimize the scan parameters is similar

Page 72

TappingMode AFMAdvanced TappingMode AFM OperationRev. D Dimension 3100 Manual 1439.4 Advanced TappingMode AFM OperationThis section discusses the more

Page 73 - Chapter 4 Installation

TappingMode AFMAdvanced TappingMode AFM Operation144 Dimension 3100 Manual Rev. Dmore rapidly. The feedback system senses the error caused by going of

Page 74 - 4.1 Shipping and Receiving

TappingMode AFMAdvanced TappingMode AFM OperationRev. D Dimension 3100 Manual 145Figure 9.4c Scope Trace with Correct Scan Rate9.4.3 Decreasing the

Page 75 - Shipping and Receiving

TappingMode AFMAdvanced TappingMode AFM Operation146 Dimension 3100 Manual Rev. D9.4.4 Optimization of Scanning ParametersThe user is encouraged to re

Page 76 - 4.2 Uncrating the System

TappingMode AFMAdvanced TappingMode AFM OperationRev. D Dimension 3100 Manual 147The Setpoint parameter defines the desired voltage for the feedback lo

Page 77

TappingMode AFMTroubleshooting148 Dimension 3100 Manual Rev. D9.5 Troubleshooting9.5.1 Frequency Response PlotIf a peak in the frequency response plot

Page 78 - Chuck Base

Rev. D Dimension 3100 Manual 149Chapter 10 Fluid ImagingThe imaging of samples under fluid is an ever-increasing realm for SPM technology. Imaging sam

Page 79 - Keyboard MouseTrackball

List of FiguresRev. D Dimension 3100 Manual xviiFigure 13.2c Tip-Sample Interaction During a Force Plot . . . . . . . . . . . . . .208Figure 13.2d A

Page 80

Fluid Imaging150 Dimension 3100 Manual Rev. D• Load the Probe: Section 10.5.3• Install the Fluid Tip Holder: Section 10.5.4• Install the Protective Sk

Page 81 - Installation

Fluid ImagingBasic PrinciplesRev. D Dimension 3100 Manual 15110.1 Basic PrinciplesAttractive forces due to surface tension effects are eliminated when

Page 82 - OK twice

Fluid ImagingFluid Operation Hardware152 Dimension 3100 Manual Rev. DFigure 10.2a Fluid Tip Holder10.2.2 Tip SuggestionsSoft cantilevers, particular

Page 83 - NanoScope® IIIa

Fluid ImagingSample MountingRev. D Dimension 3100 Manual 15310.2.3 Rubber Protective SkirtA rubber protective skirt (see Figure 10.2b) is also provide

Page 84

Fluid ImagingSample Mounting154 Dimension 3100 Manual Rev. D10.3.2 Larger SamplesFor larger samples, it may be desirable to image a small region under

Page 85 - NanoScope®

Fluid ImagingPrecautionsRev. D Dimension 3100 Manual 15510.4 Precautions10.4.1 Spillage PrecautionsThroughout all procedures outlined in the following

Page 86

Fluid ImagingPrecautions156 Dimension 3100 Manual Rev. DCAUTION: When imaging fluid samples, use extraordinary precautions against spillage. Fluids mu

Page 87

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 15710.5 Operating Principles10.5.1 Clean Fluid Cell and Protective SkirtTo reduce contam

Page 88 - 4.5 System Power-up

Fluid ImagingOperating Principles158 Dimension 3100 Manual Rev. DNote: When pushing down on the holder, the steel wire in the docking station will pre

Page 89 - Chapter 5 Stage System

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 15910.5.5 Install the Protective SkirtThe protective skirt is a rubber seal used to prot

Page 90 - 5.1 Mounting of Samples

List of Figuresxviii Dimension 3100 Manual Rev. DFigure 16.2a Comparison of Attractive and Repulsive Forces . . . . . . . . . . 275Figure 16.2b Dia

Page 91 - Mounting of Samples

Fluid ImagingOperating Principles160 Dimension 3100 Manual Rev. D10.5.7 “False” ReflectionsThe cantilever substrate rests on a smooth, angled, glass su

Page 92 - 5.2 Stage Menu Commands

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 161Remount the Dimension HeadWith the sample prepared and positioned on the stage, slowl

Page 93 - 5.2.2 Locate Tip

Fluid ImagingOperating Principles162 Dimension 3100 Manual Rev. D10.5.9 Lower Probe into FluidLower the head, allowing the fluid tip holder to enter th

Page 94 - 5.2.3 Focus Surface

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 163Note: The difference between the vertical deflection before engaging and the setpoint

Page 95

Fluid ImagingOperating Principles164 Dimension 3100 Manual Rev. D6. Adjust laser alignment in photodetectorFluid changes the optical path length as th

Page 96 - 5.2.4 Move To (X,Y)

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 165Focus Surface Procedure1. Use the Stage > Focus Surface command to bring the sampl

Page 97 - 5.2.5 Set Reference

Fluid ImagingOperating Principles166 Dimension 3100 Manual Rev. Ddamped. Instead, this step is used to find an oscillating frequency specific to the flui

Page 98 - X Axis Y AxisOrigin ResetQuit

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 1677. Center the peak on the frequency sweep plot shown on the display monitor using the

Page 99 - 5.2.6 Programmed Move

Fluid ImagingOperating Principles168 Dimension 3100 Manual Rev. D10. Adjust the Drive amplitude until obtaining an RMS Amplitude of 0.3 - 0.6V. This v

Page 100

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 16910.5.15 Show All ItemsBefore changing any parameters, you should display all of the a

Page 101

List of FiguresRev. D Dimension 3100 Manual xixFigure 17.7a Calibration Horizontal Reference . . . . . . . . . . . . . . . . . . . . . .327Figure 17.

Page 102 - Stage Menu Commands

Fluid ImagingOperating Principles170 Dimension 3100 Manual Rev. D10.5.16 Set Initial Scan ParametersScan Controls PanelIn the Scan Controls panel, set

Page 103 - 5.2.7 Initialize

Fluid ImagingOperating PrinciplesRev. D Dimension 3100 Manual 17110.5.17 Engage1. Select Motor > Engage. A pre-engage check begins, followed by Z-s

Page 104

Fluid ImagingOperating Principles172 Dimension 3100 Manual Rev. DTappingModeUsually the best images are obtained at setpoints 5-10 percent less than t

Page 105 - 5.2.8 SPM Parameters

Fluid ImagingTroubleshootingRev. D Dimension 3100 Manual 17310.6 Troubleshooting10.6.1 Cantilever Tune Plot Looks PoorBecome familiar with the charact

Page 106

Fluid ImagingTroubleshooting174 Dimension 3100 Manual Rev. D10.6.4 Unable to Locate Particulate SamplesSome particulate samples (i.e., proteins) may p

Page 107

Rev. D Dimension 3100 Manual 175Chapter 11 Scanning Tunneling Microscopy (STM)This chapter includes the following sections:• Introduction: Section 11

Page 108 - 6.1.2 Cantilever Preparation

Scanning Tunneling Microscopy (STM)Introduction176 Dimension 3100 Manual Rev. D11.1 IntroductionSTM relies on “tunneling current” between the probe an

Page 109 - Silicon Cantilever Substrates

Scanning Tunneling Microscopy (STM)Basic STM OperationRev. D Dimension 3100 Manual 177Figure 11.1a Dimension Tip Holder and Head Connection11.2 Basi

Page 110

Scanning Tunneling Microscopy (STM)Basic STM Operation178 Dimension 3100 Manual Rev. D4. Plug the loaded tip holder into the end of the Dimension head

Page 111

Scanning Tunneling Microscopy (STM)Basic STM OperationRev. D Dimension 3100 Manual 179Click on the Scope Mode command in the View menu to see what the

Page 112

Notices: The information in this document is subject to change without notice. NO WARRANTY OF ANY KIND IS MADE WITH REGARD TO THIS MATERIAL, INCLUDI

Page 113 - Press here to break out strip

List of Figuresxx Dimension 3100 Manual Rev. D

Page 114

Scanning Tunneling Microscopy (STM)STM-Specific Information and Operations180 Dimension 3100 Manual Rev. D8. After a clean image has been obtained, it

Page 115

Scanning Tunneling Microscopy (STM)STM-Specific Information and OperationsRev. D Dimension 3100 Manual 18111.3.1 STM HardwareSome individual STM compon

Page 116 - Lever Type

Scanning Tunneling Microscopy (STM)STM-Specific Information and Operations182 Dimension 3100 Manual Rev. D11.3.2 Fine Points of STM OperationGetting th

Page 117 - Scanning Profile

Scanning Tunneling Microscopy (STM)STM-Specific Information and OperationsRev. D Dimension 3100 Manual 18311.3.4 STM-Specific Menu ParametersIn addition

Page 118

Scanning Tunneling Microscopy (STM)STM-Specific Information and Operations184 Dimension 3100 Manual Rev. Dhigh will cause oscillations in the scope ima

Page 119 - Preparation

Scanning Tunneling Microscopy (STM)STM-Specific Information and OperationsRev. D Dimension 3100 Manual 185Tunneling TipsAlthough the microscope will ac

Page 120 - 7.1 System Information

Scanning Tunneling Microscopy (STM)STM-Specific Information and Operations186 Dimension 3100 Manual Rev. DVibration IsolationThe microscope should be i

Page 121 - 7.1.3 Laser Requirements

Scanning Tunneling Microscopy (STM)Spectroscopy with the STMRev. D Dimension 3100 Manual 187Head OffsetThe value of the head preamplifier offset can be

Page 122 - 7.2 Basic AFM Operation

Scanning Tunneling Microscopy (STM)Troubleshooting Operation of STM188 Dimension 3100 Manual Rev. DA comparison to STM imaging reveals two somewhat co

Page 123 - Basic AFM Operation

Scanning Tunneling Microscopy (STM)Troubleshooting Operation of STMRev. D Dimension 3100 Manual 18911.5.2 Head Engages ImmediatelyIf the STM engages i

Page 124

Rev. D Dimension 3100 Manual 1Chapter 1 System OverviewThe Dimension 3100 Scanning Probe Microscope (SPM) produces high-resolution, three-dimensiona

Page 125

Scanning Tunneling Microscopy (STM)Troubleshooting Operation of STM190 Dimension 3100 Manual Rev. Dresistance measurements. However, oxide and contami

Page 126 - 7.2.9 Align Laser

Scanning Tunneling Microscopy (STM)Troubleshooting Operation of STMRev. D Dimension 3100 Manual 191Image is Streaky or WavyIf high-resolution images a

Page 127

Scanning Tunneling Microscopy (STM)Etching Tungsten Tips192 Dimension 3100 Manual Rev. D11.6 Etching Tungsten TipsYou can purchase tungsten tips from

Page 128

Scanning Tunneling Microscopy (STM)Etching Tungsten TipsRev. D Dimension 3100 Manual 1937. Place the tungsten tips into a holder. We like to use an IC

Page 129

Scanning Tunneling Microscopy (STM)Etching Tungsten Tips194 Dimension 3100 Manual Rev. D

Page 130

Rev. D Dimension 3100 Manual 195Chapter 12 Lateral Force ModeThis chapter includes the following sections:• Introduction: Section 12.1• Basic LFM Ope

Page 131 - 7.2.10 Adjust Photodetector

Lateral Force ModeBasic LFM Operation196 Dimension 3100 Manual Rev. D12.2 Basic LFM Operation1. Set up and run the system in Contact mode as described

Page 132

Lateral Force ModeAdvanced LFM OperationRev. D Dimension 3100 Manual 19712.3 Advanced LFM Operation12.3.1 Scan DirectionThe cantilever is most suscept

Page 133 - 7.2.11 Locate Tip

Lateral Force ModeAdvanced LFM Operation198 Dimension 3100 Manual Rev. D12.3.2 Tip selectionThe analog-to-digital converter on the auxiliary input ch

Page 134 - 7.2.12 Load the Sample

Lateral Force ModeAdvanced LFM OperationRev. D Dimension 3100 Manual 199Table 12.3a Key for LFM Example Above12.3.4 Understanding the Color ScaleLF

Page 135 - 7.2.13 Focus Surface

System OverviewSystem Overview2 Dimension 3100 Manual Rev. D• Sample Size & Handling: Section 1.4• Facilities Specifications: Section 1.5• Applicat

Page 136

Lateral Force ModeAdvanced LFM Operation200 Dimension 3100 Manual Rev. D12.3.5 Using TMR Voltage to Measure FrictionThe signal called TMR (Trace minu

Page 137 - 7.3 Advanced AFM Operation

Lateral Force ModeAdvanced LFM OperationRev. D Dimension 3100 Manual 201/(198 Blank)12.3.7 Height Artifacts in the SignalLFM is subject to height art

Page 139 - Chapter 8 Contact AFM

Rev. D Dimension 3100 Manual 203Chapter 13 Force ImagingForce plots measure tip-sample interactions and determine optimal setpoints. More recently, m

Page 140 - 8.1.4 Align Laser

Force ImagingForce Plots–An Analogy204 Dimension 3100 Manual Rev. D• Force Calibration (TappingMode): Section 13.5• Force Plots: Section 13.5.1• Obtai

Page 141 - 8.1.7 Focus Surface

Force ImagingForce Plots–An AnalogyRev. D Dimension 3100 Manual 205Figure 13.1a Comparative Index of Pulling ForcesThe pulling force is measured at

Page 142 - 8.1.8 Show All Items

Force ImagingForce Calibration Mode206 Dimension 3100 Manual Rev. DThis oversimplified model depicts activity between SPM tips and various materials. I

Page 143

Force ImagingForce Calibration ModeRev. D Dimension 3100 Manual 207Figure 13.2b Piezo Travel in Force Calibration ModeAs the piezo moves the tip up

Page 144

Force ImagingForce Calibration Mode208 Dimension 3100 Manual Rev. DFigure 13.2c Tip-Sample Interaction During a Force Plot1234567

Page 145 - 8.2.1 Cantilever Selection

Force ImagingForce Calibration ModeRev. D Dimension 3100 Manual 20913.2.1 Example Force PlotA Contact Mode AFM force plot using a silicon nitride tip

Page 146 - Pull-off

System OverviewControl Station OverviewRev. D Dimension 3100 Manual 3Video Image Capture CapabilityVideo image capture capability allows the user to e

Page 147 - 8.3.1 Data Type

Force ImagingForce Calibration Mode210 Dimension 3100 Manual Rev. DFigure 13.2d Anatomy of a Force CurveThe horizontal axis plots the tip movement r

Page 148 - 8.3.2 Gain Settings

Force ImagingForce Calibration ModeRev. D Dimension 3100 Manual 211You can use force plots to adjust the setpoint so that minimal force is applied to

Page 149 - 8.3.6 Highpass Filter

Force ImagingForce Calibration Control Panels and Menus212 Dimension 3100 Manual Rev. D13.3 Force Calibration Control Panels and MenusThe Force Calibr

Page 150 - 8.4 Force Calibration Mode

Force ImagingForce Calibration Control Panels and MenusRev. D Dimension 3100 Manual 21313.3.1 Main Controls (Ramp Controls)Ramp Channel (Advanced Mode

Page 151 - Chapter 9 TappingMode AFM

Force ImagingForce Calibration Control Panels and Menus214 Dimension 3100 Manual Rev. DNumber of SamplesThis parameter defines the number of data point

Page 152 - 9.1 Principles of TappingMode

Force ImagingForce Calibration Control Panels and MenusRev. D Dimension 3100 Manual 21513.3.3 Channel 1, 2, 3 PanelsData TypeThis parameter allows you

Page 153 - Sample surface

Force ImagingForce Calibration Control Panels and Menus216 Dimension 3100 Manual Rev. D13.3.4 Feedback Controls PanelAll of the parameters in the Feed

Page 154 - 9.2.5 Locate Tip

Force ImagingForce Calibration Control Panels and MenusRev. D Dimension 3100 Manual 21713.3.5 Scan Mode Panel (Advanced Mode Only)Trigger ModeThe Scan

Page 155 - 9.2.7 Cantilever Tune

Force ImagingForce Calibration Control Panels and Menus218 Dimension 3100 Manual Rev. DTrigger ThresholdThis parameter as well as the Trigger Directio

Page 156

Force ImagingForce Calibration Control Panels and MenusRev. D Dimension 3100 Manual 219Ramp DelayThis parameter sets the amount of time to wait with t

Page 157

System OverviewControl Station Overview4 Dimension 3100 Manual Rev. D1.2.2 ComputerThe computer is the main control unit of the Dimension 3100 SPM sys

Page 158 - Frequency Sweep

Force ImagingForce Calibration (Contact Mode AFM)220 Dimension 3100 Manual Rev. DMotorThe Motor menu allows you to withdraw or manually control the ti

Page 159 - 9.2.8 Show All Items

Force ImagingForce Calibration (Contact Mode AFM)Rev. D Dimension 3100 Manual 221discusses general approach adjustments to improve force curves obtain

Page 160

Force ImagingForce Calibration (Contact Mode AFM)222 Dimension 3100 Manual Rev. D13.4.2 Helpful SuggestionsTo minimize or calculate the contact force

Page 161 - 9.2.10 Engage

Force ImagingForce Calibration (Contact Mode AFM)Rev. D Dimension 3100 Manual 223Adjust PhotodiodeIn an analogous manner, you can use the photodiode p

Page 162 - 9.3 Withdraw the Tip

Force ImagingForce Calibration (Contact Mode AFM)224 Dimension 3100 Manual Rev. DIf Sensitivity is calibrated on a material much stiffer than the cant

Page 163 - 9.4.2 Cantilever Oscillation

Force ImagingForce Calibration (Contact Mode AFM)Rev. D Dimension 3100 Manual 225Ramp size or offset the scan to a different area of the sample. Howev

Page 164 - Scan Size - 2.50 µm/div

Force ImagingForce Calibration (Contact Mode AFM)226 Dimension 3100 Manual Rev. DWhen the Data type is set to Height with the feedback gains set high,

Page 165 - Drive Frequency

Force ImagingForce Calibration (Contact Mode AFM)Rev. D Dimension 3100 Manual 22713.4.4 Interpreting Force CurvesAn examination of force curves can pr

Page 166 - 9.4.6 Gain Settings

Force ImagingForce Calibration (TappingMode)228 Dimension 3100 Manual Rev. D13.5 Force Calibration (TappingMode)Force Mode allows the imaging of force

Page 167 - 9.4.8 Surface Tune

Force ImagingForce Calibration (TappingMode)Rev. D Dimension 3100 Manual 229Figure 13.5a Piezo Extension Versus RMS Amplitude and DeflectionFigure 13

Page 168 - 9.5 Troubleshooting

System OverviewControl Station OverviewRev. D Dimension 3100 Manual 51.2.3 NanoScope IIIa ControllerThe NanoScope IIIa controller controls the microsc

Page 169 - Chapter 10 Fluid Imaging

Force ImagingForce Calibration (TappingMode)230 Dimension 3100 Manual Rev. Dthe sample surface, the oscillation amplitude of the cantilever decreases

Page 170

Force ImagingForce Calibration (TappingMode)Rev. D Dimension 3100 Manual 231Note: Collectively, these panels control tip-sample interactions. If any p

Page 171 - 10.2 Fluid Operation Hardware

Force ImagingForce Calibration (TappingMode)232 Dimension 3100 Manual Rev. D13.5.3 High Contact ForceFigure 13.5c shows a curve produced when the tip

Page 172 - 10.2.2 Tip Suggestions

Force ImagingForce ModulationRev. D Dimension 3100 Manual 23313.6 Force Modulation13.6.1 IntroductionThis section describes the operation of force mod

Page 173 - 10.3 Sample Mounting

Force ImagingForce Modulation234 Dimension 3100 Manual Rev. DFigure 13.6b Force Modulation Cantilever Holder13.6.2 Selecting a Force Modulation TipT

Page 174 - 10.3.3 Smaller Samples

Force ImagingForce ModulationRev. D Dimension 3100 Manual 235Table 13.6a Force Modulation Tips13.6.3 Operating PrincipleForce modulation mode is ver

Page 175 - 10.4 Precautions

Force ImagingForce Modulation236 Dimension 3100 Manual Rev. D4. Install the cantilever holder on the Dimension SPM head.5. Align the laser on the cant

Page 176 - Precautions

Force ImagingForce ModulationRev. D Dimension 3100 Manual 237Figure 13.6c Auto Tune Controls Paneld. Set the Drive frequency to 15kHz and the Sweep

Page 177 - 10.5 Operating Principles

Force ImagingForce Modulation238 Dimension 3100 Manual Rev. Df. Adjust the Sweep width, Data scale and the Drive amplitude until you can clearly see p

Page 178 - Operating Principles

Force ImagingForce ModulationRev. D Dimension 3100 Manual 239Note: You may also change the Drive Frequency by clicking on the Drive Frequency paramete

Page 179 - 10.5.6 Align Laser

System OverviewControl Station Overview6 Dimension 3100 Manual Rev. D1.2.4 Dimension 3100 ControllerThe Dimension 3100 controller controls the vacuum

Page 180 - 10.5.8 Load Sample

Force ImagingForce Modulation240 Dimension 3100 Manual Rev. D18. Optimize Drive amplitude for Force Modulation imaging. The amount of contrast and the

Page 181

Force ImagingForce ModulationRev. D Dimension 3100 Manual 24113.6.5 Notes About ArtifactsIt is possible to see artifacts in force modulation images th

Page 182 - 10.5.11 Adjust Photodetector

Force ImagingForce Modulation242 Dimension 3100 Manual Rev. DFigure 13.6f Friction on Force Modulation ImagesTip ShapeThe amount of indentation into

Page 183 - 10.5.13 Focus Surface

Force ImagingForce Modulation with ‘Negative LiftMode’Rev. D Dimension 3100 Manual 24313.7 Force Modulation with ‘Negative LiftMode’A new form of forc

Page 184

Force ImagingForce Modulation with ‘Negative LiftMode’244 Dimension 3100 Manual Rev. D13.7.2 Obtain a TappingMode ImageWhile negative LiftMode force m

Page 185

Force ImagingForce VolumeRev. D Dimension 3100 Manual 245/(242 Blank)Note: If the Lift scan height is too large, the interleave may be tapping, and th

Page 187

Rev. D Dimension 3100 Manual 247Chapter 14 Interleave ScanningThis chapter provides general information regarding the Interleave procedure and comman

Page 188

Interleave ScanningPreface: Interleave Scanning & LiftMode248 Dimension 3100 Manual Rev. D14.1 Preface: Interleave Scanning & LiftModeInterlea

Page 189 - 10.5.15 Show All Items

Interleave ScanningInterleave Mode DescriptionRev. D Dimension 3100 Manual 24914.2 Interleave Mode DescriptionEnabling Interleave changes the scan pat

Page 190

System OverviewDimension 3100 SPM OverviewRev. D Dimension 3100 Manual 7The Dimension 3100 controller houses the following components:Power SupplyThe

Page 191 - 10.5.17 Engage

Interleave ScanningLiftMode Description250 Dimension 3100 Manual Rev. D14.3 LiftMode DescriptionWith the Interleave scan option set to Lift, the motio

Page 192

Interleave ScanningUse of LiftMode with TappingModeRev. D Dimension 3100 Manual 251Note that certain constraints are imposed: scan sizes, offsets, ang

Page 193 - 10.6 Troubleshooting

Interleave ScanningUse of LiftMode with TappingMode252 Dimension 3100 Manual Rev. D14.5.2 Setpoint SelectionWhen the main and interleave Drive amplitu

Page 194 - Troubleshooting

Interleave ScanningUse of LiftMode with TappingModeRev. D Dimension 3100 Manual 253/(246 Blank)14.5.4 Amplitude Data InterpretationWhen monitoring amp

Page 196 - 11.1 Introduction

Rev. D Dimension 3100 Manual 255Chapter 15 Magnetic Force MicroscopyThis chapter describes how to perform Magnetic Force Microscopy (MFM) using the I

Page 197 - 11.2 Basic STM Operation

Magnetic Force MicroscopyMagnetic Force Microscopy256 Dimension 3100 Manual Rev. D• Troubleshooting: Section 15.6• MFM Image Verification: Section 15.6

Page 198 - Basic STM Operation

Magnetic Force MicroscopyMagnetic Force MicroscopyRev. D Dimension 3100 Manual 25715.1.2 Amplitude Detection TechniquesAll standard Dimension-series S

Page 199

Magnetic Force MicroscopyBasic MFM Operation258 Dimension 3100 Manual Rev. D15.2 Basic MFM OperationIn the instructions below, steps specific to phase

Page 200 - parameters

Magnetic Force MicroscopyBasic MFM OperationRev. D Dimension 3100 Manual 25915.2.2 Magnetic Force Microscopy Procedure1. Magnetize a NanoProbe magneti

Page 201 - 11.3.1 STM Hardware

System OverviewDimension 3100 SPM Overview8 Dimension 3100 Manual Rev. DFigure 1.3a D3100 Microscope Electronics Box (rear view)Note: Dimension 3100

Page 202 - 11.3.3 STM Operating Modes

Magnetic Force MicroscopyBasic MFM Operation260 Dimension 3100 Manual Rev. DTo correctly track the cantilever phase, the Phase offset parameter must b

Page 203

Magnetic Force MicroscopyBasic MFM OperationRev. D Dimension 3100 Manual 261Figure 15.2c Cantilever Tune for Amplitude DetectionFigure 15.2d Shift

Page 204

Magnetic Force MicroscopyBasic MFM Operation262 Dimension 3100 Manual Rev. D8. Under Scan Controls, set the Scan size to 5µm and Scan rate to 1–2Hz.9.

Page 205

Magnetic Force MicroscopyBasic MFM OperationRev. D Dimension 3100 Manual 263Phase DetectionSet the Channel 2 image Data type to Phase; Z range to 3 de

Page 206 - Sensitivity (Å/V)

Magnetic Force MicroscopyAdvanced MFM Operation264 Dimension 3100 Manual Rev. DTable 15.2a Frequency Modulation Gains Initial SettingsAs with topogr

Page 207 - 11.4.1 Operation of STS

Magnetic Force MicroscopyAdvanced MFM OperationRev. D Dimension 3100 Manual 265Figure 15.3a High-resolution Magnetic Force Gradient Image15.3.2 Fine

Page 208

Magnetic Force MicroscopyAdvanced MFM Operation266 Dimension 3100 Manual Rev. Dimage. It is usually safe to increase the Drive Amplitude until the firs

Page 209

Magnetic Force MicroscopyAdvanced MFM OperationRev. D Dimension 3100 Manual 267SetpointFor the most reproducible results, it is best to use a consiste

Page 210

Magnetic Force MicroscopyInstallation of the Electronics Modules268 Dimension 3100 Manual Rev. D15.4 Installation of the Electronics Modules15.4.1 Pha

Page 211

Magnetic Force MicroscopySoftware Setup Configuration (Phase, Quadrex or NSIV)Rev. D Dimension 3100 Manual 269Important Points1. Extender-compatible mi

Page 212 - 11.6 Etching Tungsten Tips

System OverviewDimension 3100 SPM OverviewRev. D Dimension 3100 Manual 91.3.2 Optics and Motors OverviewThe optic system assists you in locating the c

Page 213 - Etching Tungsten Tips

Magnetic Force MicroscopyTroubleshooting270 Dimension 3100 Manual Rev. DNote: This step is not necessary for NanoScope IV.5. Click the Ok button when

Page 214

Rev. D Dimension 3100 Manual 271Chapter 16 Electric TechniquesThis chapter describes how to perform two electric techniques: Electric Force Microscop

Page 215 - 12.1 Introduction

Electric Techniques272 Dimension 3100 Manual Rev. D• Optimize Tune in Interleave: Section 16.3.6• If Voltage is Needed, Use Analog 2 When Possible: Se

Page 216 - 12.2 Basic LFM Operation

Electric TechniquesElectric Techniques OverviewRev. D Dimension 3100 Manual 27316.1 Electric Techniques OverviewThere are two types of electric techni

Page 217 - 12.3 Advanced LFM Operation

Electric TechniquesElectric Techniques Overview274 Dimension 3100 Manual Rev. DFigure 16.1b Extender Electronics ModuleIf applying a voltage to the

Page 218 - 12.3.2 Tip selection

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 27516.2 Electric Force Microscopy16.2.1 Electric Force Microscopy TheoryElect

Page 219 - Low Friction

Electric TechniquesElectric Force Microscopy276 Dimension 3100 Manual Rev. DFigure 16.2b Diagram of Extender Electronics Module for EFMIn many cases

Page 220 - Advanced LFM Operation

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 277Figure 16.2c Microscope BackplaneInstructions for Reconfiguring JumpersCa

Page 221

Electric TechniquesElectric Force Microscopy278 Dimension 3100 Manual Rev. DSetting Analog 2 in the SoftwareIf you have chosen a configuration where An

Page 222

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 279Setting the Extender Electronics Box1. For systems with an Extender Electr

Page 223 - Chapter 13 Force Imaging

Rev. D Dimension 3100 Manual iii Ta ble of Contents Chapter 1 System Overview 1 1.1 System Overview . . . . . . . . . . . . . . . . . . . . . . . .

Page 224 - 13.1 Force Plots–An Analogy

System OverviewDimension 3100 SPM Overview10 Dimension 3100 Manual Rev. DFigure 1.3b Dimension SPM HeadThe Dimension SPM head scans the tip and gene

Page 225 - Force Plots–An Analogy

Electric TechniquesElectric Force Microscopy280 Dimension 3100 Manual Rev. DJumper Configurations Without Extender ElectronicsAs shipped from the facto

Page 226 - 13.2 Force Calibration Mode

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 281Analog 2 Voltage Applied to the Tip (No Extender Electronics)The jumper co

Page 227 - Force Calibration Mode

Electric TechniquesElectric Force Microscopy282 Dimension 3100 Manual Rev. DAnalog 2 Voltage Applied to the Sample (No Extender Electronics)The jumper

Page 228

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 283External Voltage Source Applied to the Tip (No Extender Electronics)In som

Page 229 - 13.2.1 Example Force Plot

Electric TechniquesElectric Force Microscopy284 Dimension 3100 Manual Rev. DExternal Voltage Source Applied to Sample (No Extender Electronics)In some

Page 230

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 285Jumper Configurations With Extender ElectronicsAs shipped from the factory,

Page 231

Electric TechniquesElectric Force Microscopy286 Dimension 3100 Manual Rev. DAnalog 2 Voltage Applied to the Tip (With Extender Electronics)Notice that

Page 232

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 287Analog 2 Voltage Applied to Sample (With Extender Electronics)The jumper c

Page 233 - Y Offset

Electric TechniquesElectric Force Microscopy288 Dimension 3100 Manual Rev. DExternal Voltage Source Applied to Tip (With Extender Electronics)In some

Page 234

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 289External Voltage Source Applied to Sample (With Extender Electronics)In so

Page 235 - 13.3.3 Channel 1, 2, 3 Panels

System OverviewDimension 3100 SPM OverviewRev. D Dimension 3100 Manual 11Adjustable Detector MirrorAn adjustable mirror positions the reflected laser s

Page 236

Electric TechniquesElectric Force Microscopy290 Dimension 3100 Manual Rev. D16.2.3 Electric Force Microscopy ProceduresNote: Amplitude detection is th

Page 237 - Total Force

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 29116.2.4 Phase DetectionPhase Detection is only available when the Extender

Page 238 - End Mode

Electric TechniquesElectric Force Microscopy292 Dimension 3100 Manual Rev. D3. Quit Auto Tune.4. Select Back to Image Mode.5. Engage the AFM and make

Page 239 - 13.3.6 Menu Bar Commands

Electric TechniquesElectric Force MicroscopyRev. D Dimension 3100 Manual 29316.2.5 Frequency ModulationFor more quantitative results, use frequency mo

Page 240 - Position

Electric TechniquesElectric Force Microscopy294 Dimension 3100 Manual Rev. DFigure 16.2r Amplitude Detection Cantilever Tune2. For maximum sensitivi

Page 241

Electric TechniquesEFM Troubleshooting/PointersRev. D Dimension 3100 Manual 29516.3 EFM Troubleshooting/Pointers16.3.1 Use Low Setpoint When Tapping i

Page 242 - 13.4.2 Helpful Suggestions

Electric TechniquesSurface Potential Detection—Theory296 Dimension 3100 Manual Rev. D16.3.6 Optimize Tune in InterleaveFind optimal lift height. Use V

Page 243 - 13.4.3 Advanced Techniques

Electric TechniquesSurface Potential Detection—TheoryRev. D Dimension 3100 Manual 297oscillating electrostatic force at the frequency ω on the cantile

Page 244

Electric TechniquesSurface Potential Detection—Preparation298 Dimension 3100 Manual Rev. D16.5 Surface Potential Detection—PreparationIt is often desi

Page 245

Electric TechniquesSurface Potential Detection—PreparationRev. D Dimension 3100 Manual 299Figure 16.5a Normal Jumper Configuration with Extender Elec

Page 246

System OverviewDimension 3100 SPM Overview12 Dimension 3100 Manual Rev. DFigure 1.3d Dimension Head ScannerThe end of the scanner allows easy remova

Page 247

Electric TechniquesSurface Potential Detection—Preparation300 Dimension 3100 Manual Rev. DAnalog 2 Voltage Applied to Sample (With Extender Electronic

Page 248 - 13.5.1 Force Plots

Electric TechniquesSurface Potential Imaging—ProcedureRev. D Dimension 3100 Manual 30116.6 Surface Potential Imaging—Procedure1. Locate the two toggle

Page 249

Electric TechniquesSurface Potential Imaging—Procedure302 Dimension 3100 Manual Rev. DVerify that all deposited metal adheres strongly to the silicon

Page 250

Electric TechniquesSurface Potential Imaging—ProcedureRev. D Dimension 3100 Manual 303amplitude at zero. To do this, the feedback circuit uses the loc

Page 251

Electric TechniquesSurface Potential Imaging—Procedure304 Dimension 3100 Manual Rev. D“Potential” signal), it is possible to use the lock-in signal to

Page 252 - 13.5.4 Tip Selection

Rev. D Dimension 3100 Manual 305Chapter 17 CalibrationThis chapter provides detailed instructions for the fine calibration of Digital Instruments Dime

Page 253 - 13.6 Force Modulation

Calibration306 Dimension 3100 Manual Rev. D• Fine-tuning for X-Y Calibration: Section 17.7• Prepare System for Fine-tuning: Section 17.7.1• Measure Ho

Page 254 - Force Modulation

CalibrationSPM Calibration OverviewRev. D Dimension 3100 Manual 30717.1 SPM Calibration OverviewVeeco employs a software-guided calibration procedure

Page 255 - 13.6.3 Operating Principle

CalibrationSPM Calibration Overview308 Dimension 3100 Manual Rev. DTable 17.1a Calibration ScheduleNote: Small Scan Size Calibration: If using scan

Page 256

CalibrationSPM Calibration OverviewRev. D Dimension 3100 Manual 309Figure 17.1a Scanner Crystal Voltage and Photodiode VoltageThe Microscope > Ca

Page 257

System OverviewDimension 3100 SPM OverviewRev. D Dimension 3100 Manual 131.3.5 Cantilever HolderThe cantilever holder is a small printed circuit card

Page 258

CalibrationSPM Calibration Overview310 Dimension 3100 Manual Rev. DFigure 17.1b Scanner Sensitivity CurveThis curve typifies scanner sensitivity acro

Page 259

CalibrationSPM Calibration OverviewRev. D Dimension 3100 Manual 311Figure 17.1d Scanner Voltage and MovementThrough rigorous quality control of its

Page 260

CalibrationCalibration Setup312 Dimension 3100 Manual Rev. D17.1.2 Calibration ReferencesAs described above, each scanner exhibits its own unique sens

Page 261 - 13.6.5 Notes About Artifacts

CalibrationCalibration SetupRev. D Dimension 3100 Manual 31317.2.2 Align Calibration Reference1. Load the silicon calibration reference into the SPM.2

Page 262

CalibrationCheck Sample Orthogonality314 Dimension 3100 Manual Rev. D17.2.4 Set Up Contact AFM1. Set the AFM mode to Contact.Note: The microscope can

Page 263

CalibrationCheck Sample OrthogonalityRev. D Dimension 3100 Manual 315Note: In Figure 17.3a, pits align with the vertical (slow) axis but skew with the

Page 264

CalibrationLinearity Correction316 Dimension 3100 Manual Rev. D17.4 Linearity CorrectionFor applications which demand good linearity, the following pr

Page 265 - 13.8 Force Volume

CalibrationLinearity CorrectionRev. D Dimension 3100 Manual 3179. Align one side of the zoom box with desired features and observe how the other side

Page 266

CalibrationLinearity Correction318 Dimension 3100 Manual Rev. D5. After setting Fast mag0 and Fast arg, insert the values for Slow mag0 and Slow arg.

Page 267

CalibrationLinearity CorrectionRev. D Dimension 3100 Manual 319Adjusting Slow arg 1. Follow the same instructions for Fast arg.2. Adjust the zoom box

Page 268

System OverviewDimension 3100 SPM Overview14 Dimension 3100 Manual Rev. DTable 1.3b Cantilever Holder SpecificationsFigure 1.3f Fluid Cell Cantile

Page 269 - Interleave Mode Description

CalibrationLinearity Correction320 Dimension 3100 Manual Rev. DFine AdjustmentInitial adjustment is usually adequate; however, if more precision is de

Page 270 - 14.3 LiftMode Description

CalibrationX-Y Calibration using Capture Calibration and AutocalibrationRev. D Dimension 3100 Manual 32117.5 X-Y Calibration using Capture Calibration

Page 271

CalibrationX-Y Calibration using Capture Calibration and Autocalibration322 Dimension 3100 Manual Rev. DFigure 17.5b Capture Control PromptNote: The

Page 272 - 14.5.2 Setpoint Selection

CalibrationAutocalibrationRev. D Dimension 3100 Manual 3239. Go to the Capture directory where all Capture Calibration files are saved.10. Select Offl

Page 273

CalibrationAutocalibration324 Dimension 3100 Manual Rev. D4. For normal calibration, verify that all parameters are selected in the dialog box, then c

Page 274

CalibrationAutocalibrationRev. D Dimension 3100 Manual 325due to the unusual movements employed during Capture Calibration scanning. Regardless, featu

Page 275

CalibrationFine-tuning for X-Y Calibration326 Dimension 3100 Manual Rev. D17.7 Fine-tuning for X-Y CalibrationFine-tuning is usually performed at two

Page 276 - Magnetic Force Microscopy

CalibrationFine-tuning for X-Y CalibrationRev. D Dimension 3100 Manual 327Figure 17.7a Calibration Horizontal Reference5. Verify that the microscope

Page 277

CalibrationFine-tuning for X-Y Calibration328 Dimension 3100 Manual Rev. D8. Multiply the quotient obtained above by the X fast sens value shown on th

Page 278 - 15.2 Basic MFM Operation

CalibrationFine-tuning for X-Y CalibrationRev. D Dimension 3100 Manual 32917.7.4 Measure Horizontally at 150 V Scan Size1. Verify that the Scan angle

Page 279 - Basic MFM Operation

System OverviewSample Size & HandlingRev. D Dimension 3100 Manual 15• Through-the-lens illumination• Color video camera •Focus tracking and automa

Page 280

CalibrationFine-tuning for X-Y Calibration330 Dimension 3100 Manual Rev. D4. Return to the Scanner Calibration dialog box.5. Select the X fast derate

Page 281

CalibrationCalibrating ZRev. D Dimension 3100 Manual 33117.8 Calibrating ZIn terms of obtaining accurate Z-axis measurements, it is generally not diffi

Page 282

CalibrationCalibrating Z332 Dimension 3100 Manual Rev. D6. If the Z Center Position value is not close to zero, use the Realtime > Motor > Tip U

Page 283

CalibrationCalibrating ZRev. D Dimension 3100 Manual 33317.8.3 Measure Vertical FeaturesWith the image corrected, its vertical features may now be mea

Page 284 - 15.3 Advanced MFM Operation

CalibrationCalibrating Z334 Dimension 3100 Manual Rev. D4. If the two peaks do not appear in the display, increase the Histogram filter cut off in the

Page 285 - 15.3.3 Drive Amplitude

CalibrationCalibrating ZRev. D Dimension 3100 Manual 33517.8.4 Correct Z SensitivityIf the depth of the pit on the 10µm silicon calibration reference

Page 286 - Advanced MFM Operation

CalibrationCalibrating Z336 Dimension 3100 Manual Rev. D4. Record the measured depth. If the depth measured by the extended piezo is off by more than

Page 287

Rev. D Dimension 3100 Manual 337Chapter 18 Maintenance, Troubleshooting and WarrantyThis chapter includes the following sections:• Maintenance: Secti

Page 288 - 15.4.1 Phase Extender Module

Maintenance, Troubleshooting and WarrantyMaintenance338 Dimension 3100 Manual Rev. D18.1 MaintenanceGenerally, the Dimension 3100 system requires very

Page 289 - 15.4.3 NanoScope IV

Maintenance, Troubleshooting and WarrantyMaintenanceRev. D Dimension 3100 Manual 3396. Wipe the top and bottom of the manifold.7. Reconnect the manifo

Page 290 - 15.6 Troubleshooting

System OverviewFacilities Specifications16 Dimension 3100 Manual Rev. D1.5 Facilities SpecificationsCompliance with the requirements and specifications o

Page 291

Maintenance, Troubleshooting and WarrantyMaintenance340 Dimension 3100 Manual Rev. D18.1.2 CalibrationA description of the complete scanner calibratio

Page 292

Maintenance, Troubleshooting and WarrantyMaintenanceRev. D Dimension 3100 Manual 34118.1.3 Changing the Illuminator Light BulbYou can change the illum

Page 293 - Electric Techniques Overview

Maintenance, Troubleshooting and WarrantyMaintenance342 Dimension 3100 Manual Rev. D5. Remove the used bulb by unplugging the bulb cable and lifting t

Page 294

Maintenance, Troubleshooting and WarrantyMaintenanceRev. D Dimension 3100 Manual 34318.1.4 Fuse Characteristics and Replacement• NanoScope Controller:

Page 295 - Electric Force Microscopy

Maintenance, Troubleshooting and WarrantyMaintenance344 Dimension 3100 Manual Rev. D3. Using a screwdriver, press the slot in the fuse holder and rota

Page 296

Maintenance, Troubleshooting and WarrantyMaintenanceRev. D Dimension 3100 Manual 34518.1.8 Optics Zoom SystemThe clutch in the zoom system can begin t

Page 297

Maintenance, Troubleshooting and WarrantyTroubleshooting346 Dimension 3100 Manual Rev. D18.2 TroubleshootingThe Dimension 3100 system is very reliable

Page 298

Maintenance, Troubleshooting and WarrantyTroubleshootingRev. D Dimension 3100 Manual 347•“Illuminator Error, Manual override is on” – The manual contr

Page 299 - NanoScope

Maintenance, Troubleshooting and WarrantyTroubleshooting348 Dimension 3100 Manual Rev. D18.2.2 Common ProblemsBent stage flexures – The X or Y stage fle

Page 300

Maintenance, Troubleshooting and WarrantyWarrantyRev. D Dimension 3100 Manual 349The scanner plug is very durable and typically does not fail due to u

Page 301

Rev. D Dimension 3100 Manual 17Chapter 2 SafetyThis chapter details the safety requirements involved in installation of the Dimension 3100 Scanning

Page 302

Maintenance, Troubleshooting and WarrantyWarranty350 Dimension 3100 Manual Rev. D Warranty EligibilityTo be eligible for the above warranties, purchas

Page 303

Rev. D Dimension 3100 Manual 351 IndexSymbols 12. 248AAirDimension 3100 controller 7Aliasing 241Amplitude 146Atomic Force Microscope (AFM)operator pre

Page 304

Index352 Dimension 3100 Manual Rev. DDeflection 127Differential Signals 11Dimension 3100 Controllerair pump 7illumination 7power supply 7system overvi

Page 305

IndexRev. D Dimension 3100 Manual 353safety 19thermal hazard 19IlluminationDimension 3100 controller 7Input and Display Devicescomputer 4keyboard 3mon

Page 306

Index354 Dimension 3100 Manual Rev. Dsystem overview 5Number of samples 214OObjectiveoptics and motors overview 9video zoom 14Offset 136, 167Operator

Page 307

IndexRev. D Dimension 3100 Manual 355precautions 26Samplesbiological 149—174chucks 70handling 15minimizing surface forces 224preparation 70, 114—115,

Page 308

Index356 Dimension 3100 Manual Rev. DThermal Hazardsymbol 19Tip crash 118Tip Down 220, 222Tip Holder 10fluid 151installation fixture 104preparation 10

Page 309

Safety18 Dimension 3100 Manual Rev. D• Software Power-up: Section 2.8• Log into Windows NT: Section 2.8.1• Log On: Section 2.8.2• Start the NanoScope

Page 310

SafetySafety RequirementsRev. D Dimension 3100 Manual 192.1 Safety RequirementsFigure 2.1a Safety Symbols KeySymbol DefinitionThis symbol identifies c

Page 311 - 16.2.4 Phase Detection

iv Dimension 3100 Manual Rev. D 2.6.2 Turn on the Dimension 3100 SPM (Service and Installation Only) . . . . . . . 30 2.6.3 Power-up Checklist (Se

Page 312

SafetySafety Precautions20 Dimension 3100 Manual Rev. D2.2 Safety PrecautionsBecause the Dimension 3100 SPM features independently motorized component

Page 313 - 16.2.6 Amplitude Detection

SafetySafety PrecautionsRev. D Dimension 3100 Manual 21CAUTION: Please contact Veeco before attempting to move the Dimension 3100 SPM system.ATTENTION

Page 314

SafetySafety Precautions22 Dimension 3100 Manual Rev. DWARNING: The Dimension 3100 SPM contains a diode laser with an output of less than 1.0mW at 670

Page 315 - EFM Troubleshooting/Pointers

SafetySafety PrecautionsRev. D Dimension 3100 Manual 232.2.2 MicroscopeTo avoid operator injury and equipment damage, observe the following cautions r

Page 316 - 16.3.8 Try Uncoated Si Tip

SafetySafety Precautions24 Dimension 3100 Manual Rev. DCAUTION: Stage microscopes feature an automated X-Y stage and Z-axis capable of programmed move

Page 317

SafetySafety PrecautionsRev. D Dimension 3100 Manual 25WARNING: Do not attempt repairs on electrical components. If it is necessary to enter the elect

Page 318

SafetyErgonomics26 Dimension 3100 Manual Rev. D2.2.3 Sample Safeguards2.3 ErgonomicsThe Dimension 3100 SPM design promotes compatibility in the integr

Page 319

SafetyEnvironmental ConditionsRev. D Dimension 3100 Manual 272.4 Environmental ConditionsEnvironmental conditions that may affect the performance of t

Page 320

SafetyPower-up Sequence (Installation and Service Only)28 Dimension 3100 Manual Rev. D2.6.1 Pre Power-up Checklist Pre-Installation___

Page 321

SafetyPower-up Sequence (Installation and Service Only)Rev. D Dimension 3100 Manual 29______ 4. Install the control station by completing the followin

Page 322

Rev. D Dimension 3100 Manual v Chapter 5 Stage System 69 5.1 Mounting of Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 323

SafetyPower-up Sequence (Installation and Service Only)30 Dimension 3100 Manual Rev. D______ 25-pin D cable from computer to NanoScope controller_____

Page 324

SafetyPower-Up Sequence (Normal Usage)Rev. D Dimension 3100 Manual 312.6.3 Power-up Checklist (Service and Installation Only)Power-up (Installation On

Page 325 - Chapter 17 Calibration

SafetySoftware Power-up32 Dimension 3100 Manual Rev. D2.7.2 Power-up Checklist (Normal Usage)Power-up (Normal Usage)_______ 1. Verify that all system

Page 326 - Section 17.8.6

SafetySoftware Power-upRev. D Dimension 3100 Manual 33Figure 2.8a Log into Windows NT2.8.2 Log OnIn the Logon Information window, enter the default

Page 327 - 17.1 SPM Calibration Overview

SafetySoftware Power-up34 Dimension 3100 Manual Rev. DFigure 2.8b Logon Window2.8.3 Start the NanoScope SoftwareGo to the desktop and click on the N

Page 328

SafetyHazard LabelsRev. D Dimension 3100 Manual 352.8.5 Begin Stage Initialization• Select Stage > Initialize.Note: Various axes of motion move to

Page 329 - SPM Calibration Overview

SafetyHazard Labels36 Dimension 3100 Manual Rev. D2.9.1 Laser Warning LabelsLaser Explanatory LabelThe Laser Explanatory Label (see Figure 2.9a) indic

Page 330

Rev. D Dimension 3100 Manual 37Chapter 3 Facilities RequirementsThis chapter details facility site requirements, safety requirements, and configurati

Page 331

Facilities RequirementsOptional Configurations38 Dimension 3100 Manual Rev. D3.1 Optional ConfigurationsThe following are typical configurations for the

Page 332 - 17.2 Calibration Setup

Facilities RequirementsOptional ConfigurationsRev. D Dimension 3100 Manual 393.1.2 VT-102This configuration consists of two basic elements: the VT-102 w

Page 333 - Scan Angle = 0°

vi Dimension 3100 Manual Rev. D7.3 Advanced AFM Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1177.3.1 Stage Param

Page 334 - 17.3.1 Measure Orthogonality

Facilities RequirementsOptional Configurations40 Dimension 3100 Manual Rev. D3.1.3 IS3K-2This configuration is comprised of one basic element, the IS3K-

Page 335 - Check Sample Orthogonality

Facilities RequirementsFacilities RequirementsRev. D Dimension 3100 Manual 413.2 Facilities RequirementsFigure 3.2a Dimension 3100 SPM Facility Requ

Page 336 - 17.4 Linearity Correction

Facilities RequirementsAcoustic/Vibration Isolation Systems42 Dimension 3100 Manual Rev. D3.3 Acoustic/Vibration Isolation Systems3.3.1 IS3K-2 Dimensi

Page 337 - Linearity Correction

Facilities RequirementsAcoustic/Vibration Isolation SystemsRev. D Dimension 3100 Manual 43Figure 3.3b IS3K-2 - Side ViewFigure 3.3c IS3K-2 - Top V

Page 338

Facilities RequirementsAcoustic/Vibration Isolation Systems44 Dimension 3100 Manual Rev. DFigure 3.3d IS3K-2 Leveling Feet Location - Bottom ViewFig

Page 339 - 17.4.2 Adjust Fast mag1

Facilities RequirementsAcoustic/Vibration Isolation SystemsRev. D Dimension 3100 Manual 453.3.2 VT-103-3K Dimensions, Utilities and ClearanceAn isolat

Page 340 - 17.4.3 Adjust Slow mag1

Facilities RequirementsAcoustic/Vibration Isolation Systems46 Dimension 3100 Manual Rev. DFigure 3.3h VT-103-3K - Top View with Acoustic Hood Closed

Page 341 - Autocalibration

Facilities RequirementsAcoustic/Vibration Isolation SystemsRev. D Dimension 3100 Manual 473.3.4 Computer/Controller Facility RequirementsThe IS3K-2 al

Page 342 - SKIP to

Facilities RequirementsAcoustic/Vibration Isolation Systems48 Dimension 3100 Manual Rev. D3.3.5 ELCON ConsoleThe ELCON Console is available as an opti

Page 343 - 17.6 Autocalibration

Facilities RequirementsFacilities Requirements SummaryRev. D Dimension 3100 Manual 493.4 Facilities Requirements SummaryIf the Dimension 3100 system i

Page 344

Rev. D Dimension 3100 Manual vii9.4.5 Data Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 345

Facilities RequirementsAcoustic/Vibration Specifications50 Dimension 3100 Manual Rev. D3.5 Acoustic/Vibration SpecificationsThe following conditions mus

Page 346

Facilities RequirementsGeneral Facilities GuidelinesRev. D Dimension 3100 Manual 51/(54 Blank)3.6 General Facilities GuidelinesThe following list cont

Page 348

Rev. D Dimension 3100 Manual 53Chapter 4 InstallationThis chapter details the installation procedure for the Dimension 3100 Scanning Probe Microscop

Page 349

InstallationShipping and Receiving54 Dimension 3100 Manual Rev. D4.1 Shipping and Receiving4.1.1 Equipment RequirementsThe following equipment is nece

Page 350

InstallationShipping and ReceivingRev. D Dimension 3100 Manual 55Cables Received• Controller-to-Dimension 3100 Cable, 37-pin D• Dimension 3100 DC Powe

Page 351 - 17.8 Calibrating Z

InstallationUncrating the System56 Dimension 3100 Manual Rev. D4.2 Uncrating the System4.2.1 Uncrate the Dimension 3100 SPM System1. Using scissors, c

Page 352 - Calibrating Z

InstallationInstalling the Dimension 3100 SystemRev. D Dimension 3100 Manual 574.3 Installing the Dimension 3100 System4.3.1 Install the Dimension 310

Page 353

InstallationInstalling the Dimension 3100 System58 Dimension 3100 Manual Rev. D8. Align the two sets of three mounting holes on the long edges of the

Page 354

InstallationInstalling the Dimension 3100 SystemRev. D Dimension 3100 Manual 594.3.2 Install the Control Station1. Set up a table to be used as the co

Page 355 - 17.8.4 Correct Z Sensitivity

viii Dimension 3100 Manual Rev. DChapter 11 Scanning Tunneling Microscopy (STM) 17511.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . .

Page 356

InstallationConnecting the Dimension 3100 System60 Dimension 3100 Manual Rev. D4.4 Connecting the Dimension 3100 System4.4.1 Dimension 3100 Control St

Page 357

InstallationConnecting the Dimension 3100 SystemRev. D Dimension 3100 Manual 61Figure 4.4a Computer (rear view)02912356Video 25Video 18USB port79COM

Page 358 - 18.1 Maintenance

InstallationConnecting the Dimension 3100 System62 Dimension 3100 Manual Rev. DNote: The computer ships with the network board disabled to avoid error

Page 359 - Maintenance

InstallationConnecting the Dimension 3100 SystemRev. D Dimension 3100 Manual 63Connect the NanoScope ControllerNote: The NanoScope III and IIIa are ad

Page 360 - 18.1.2 Calibration

InstallationConnecting the Dimension 3100 System64 Dimension 3100 Manual Rev. DConnect the Dimension 3100 Controller1. Connect extensions to the Dimen

Page 361

InstallationConnecting the Dimension 3100 SystemRev. D Dimension 3100 Manual 65Figure 4.4e Dimension 3100 Controller (front view)2. To set illuminat

Page 362

InstallationConnecting the Dimension 3100 System66 Dimension 3100 Manual Rev. D4.4.2 Dimension 3100 Microscope External Components ConnectionsDimensio

Page 363

InstallationConnecting the Dimension 3100 SystemRev. D Dimension 3100 Manual 672. Toggle the vacuum power switch, located on the front of the microsco

Page 364 - 18.1.7 Air Table Feet

InstallationSystem Power-up68 Dimension 3100 Manual Rev. D4. Disconnect all cables from the Dimension 3100 microscope, and remove the cables from the

Page 365 - 18.1.8 Optics Zoom System

Rev. D Dimension 3100 Manual 69Chapter 5 Stage SystemThe Dimension 3100 Scanning Probe Microscope (SPM) features a large sample stage capable of pos

Page 366 - 18.2 Troubleshooting

Rev. D Dimension 3100 Manual ix13.4.2 Helpful Suggestions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22213.4.

Page 367

Stage SystemMounting of Samples70 Dimension 3100 Manual Rev. D5.1 Mounting of SamplesThere are two methods widely used for mounting samples: vacuum ch

Page 368 - 18.2.2 Common Problems

Stage SystemMounting of SamplesRev. D Dimension 3100 Manual 715.1.3 Axis Orientation—Motorized X-Y StagesWhen viewing the Dimension 3100 SPM from the

Page 369 - 18.3 Warranty

Stage SystemStage Menu Commands72 Dimension 3100 Manual Rev. D5.2 Stage Menu CommandsImportant stage menu commands are discussed in detail in the foll

Page 370 - Warranty

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 732. To load a new sample, click OK. The head raises to the Load/Unload height, and the st

Page 371

Stage SystemStage Menu Commands74 Dimension 3100 Manual Rev. DFigure 5.2c Locate Tip Prompt2. Bring the tip into focus by holding down the left trac

Page 372

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 75Figure 5.2d Focus Surface Prompt2. Use the left button on the trackball to focus on th

Page 373

Stage SystemStage Menu Commands76 Dimension 3100 Manual Rev. D5.2.4 Move To (X,Y)This option enables the operator to quickly index the stage to a defin

Page 374 - 136, 167

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 775. To exit the Move To (x,y) dialog box, click the Quit button.5.2.5 Set ReferenceThe Se

Page 375

Stage SystemStage Menu Commands78 Dimension 3100 Manual Rev. DFigure 5.2g Set Reference Prompt3. Click Origin in the Set Reference panel; this is th

Page 376

Stage SystemStage Menu CommandsRev. D Dimension 3100 Manual 79Figure 5.2i Resultant Reference Line5.2.6 Programmed MoveThe Programmed Move function