Head, Probe, & Sample Preparation
Basic AFM Operation
Rev. D Dimension 3100 Manual 113
Figure 7.2j Photodetector Mirror Adjustment Knobs
7.2.11 Locate Tip
For initial start-up where the Z-stage and optics have been almost fully retracted (such as when you
performed Stage > Load New Sample), use the Focus Surface command to move the tip closer to
the surface. It is easier to focus on the tip with more light reflecting from the surface into the optics.
To locate the tip, complete the following steps:
1. Initiate the Locate Tip command in the Stage menu by clicking on Locate Tip.
Note: This command locates the tip position (Z height) using optical focal distance
measurements. When this step completes, the computer records the tip position
in memory.
2. Zoom out as far as possible using the trackball or Zoom Out button. This helps you locate
the cantilever in optical field of view.
3. Once you find the tip, use the optic adjustment knobs (lower-left corner of the zoom optics
assembly) to align the optical microscope lens so the tip is centered in the video display
window.
4. Focus on the cantilever using the trackball. You can zoom in using the Zoom button on the
trackball.
5. Quit the Locate Tip sequence by pressing ENTER on the keyboard or clicking OK at the
bottom of the Locate Tip dialog box.
(Vertical)
AFM / TM Detector
Mirror Adjustment
(Horizontal)
LFM Detector
Mirror Adjustment